Total endovascular aortic arch repair: a comprehensive review comparing parallel graft techniques and custom-made devices

Patrícia Carvalho1  2 

Daniel Brandão1  2 

Andreia Coelho1  2 

Luís Fernandes1  2 

Marta Machado1  2 

Francisco Basílio1  2 

Beatriz Guimarães1 

Ana Margarida Rocha1 

Alexandra Canedo1  2 

1 Department of Angiology and Vascular Surgery, Unidade Local de Saúde Gaia e Espinho, Portugal

2 Faculty of Medicine, University of Porto, Portugal


Abstract

Introduction:

Endovascular approaches have emerged as promising alternatives for treating aortic arch pathology, particularly in high-risk patients. This review provides a comprehensive analysis of total endovascular aortic arch reconstruction, focusing on Parallel Graft Techniques (PGTs) and Custom-Made Devices (CMDs).

Methods:

PubMed and Scopus were searched to conduct a comprehensive review comparing the technical aspects, clinical outcomes, strengths, and limitations of PGTs and CMDs. Only studies reporting on total endovascular aortic aneurysm repair were considered. Studies with fewer than five patients, published in a language other than Portuguese, English, or Spanish, and reporting hybrid/open procedures or partial aortic arch reconstructions were excluded. Findings were summarised descriptively.

Results:

A total of 38 studies comprising 871 patients were included (214 in PGT and 657 in the CMD group). Technical success rates were high in both groups, mostly ranging from 80% to 100%. No clear superiority was observed between the two approaches for thirty-day, aorta-related, or overall mortality. However, stroke and spinal cord ischaemia (SCI) rates appeared higher in the CMD group, ranging from 0 to 42.9% and 0 to 9.1%, respectively. The occurrence of type I and III endoleaks was higher in the PGT group (0-45.5% versus 0-32.4%), though this did not translate into a clear difference in reintervention rates. During follow-up, high target vessel patency rates were observed in both groups, typically exceeding 95%.

Conclusion:

PGTs appear to have non-inferior outcomes in terms of stroke rate and in-hospital and overall mortality. Although the occurrence of type I/III endoleaks remains higher in PGTs, this does not seem to affect reintervention rates. Thus, PGTs should be considered a viable option for treating complex aortic arch pathology, particularly in fragile patients, those with adverse anatomy, or in emergent situations. Large-volume prospective studies directly comparing these two techniques are currently warranted.

Keywords: Endovascular Aneurysm Repair (MeSH); Aorta, Thoracic (MeSH); Aortic Arch Syndromes (MeSH); Stroke (MeSH); Endoleak (MeSH)

Introduction

The management of aortic arch pathology is often demanding and poses significant challenges because of its anatomical complexity, high-flow dynamics, and potential involvement of the supra-aortic trunks. The history of aortic arch repair has evolved from open surgery to hybrid debranching procedures and, ultimately, to total endovascular strategies. Open surgical repair, though traditionally considered the gold standard, carries substantial morbidity and mortality, particularly in elderly patients with multiple comorbidities.

In this context, Thoracic Endovascular Aortic Repair (TEVAR) has emerged as a minimally invasive alternative, particularly in surgically unfit patients.1,2 Nevertheless, its application in the aortic arch requires not only an adequate proximal landing zone but also preservation of supra-aortic branch perfusion, which has driven the development of several endovascular techniques, such as Parallel Graft Techniques (PGTs) and branched or fenestrated Custom-Made Devices (CMDs).3,4 In fact, such endovascular strategies offer several advantages, including shorter procedural times, reduced perioperative complications, and faster recovery.

Despite the increasing interest in this field, current evidence remains limited to small observational studies with a paucity of robust comparative data regarding these endovascular approaches, particularly when applied to TEVAR with proximal landing in Ishimaru’s zone 0.

This narrative review aims to provide a comprehensive comparative analysis of total endovascular aortic arch reconstruction, focusing on PGTs and CMDs, their technical aspects, clinical outcomes, strengths and limitations.

Methods

A comprehensive search was conducted in the PubMed and Scopus databases using the following query: (endovascular aortic repair OR (chimney OR branched OR fenestrated)) AND (aortic arch pathology OR aortic arch aneurysm OR aortic arch pseudoaneurysm OR aortic arch dissection OR aortic arch penetrating ulcer OR Ishimaru zone 0). Study selection followed predefined inclusion criteria: articles published in the past ten years on total endovascular aortic arch repair with proximal landing in Ishimaru’s zone 0, using PGTs or CMDs to preserve both the Brachiocephalic Trunk (BCT) and the Left Common Carotid Artery (LCCA). Studies with fewer than five patients, published in a language other than Portuguese, English or Spanish, or reporting hybrid/open procedures or partial aortic arch reconstructions were excluded. Given the narrative nature of this review, findings were summarised descriptively rather than by statistical pooling.

Results

Technical considerations of parallel graft techniques

The PGTs represent off-the-shelf solutions for endovascular aortic arch repair, allowing preservation of blood flow to the BCT, LCCA, and/or left subclavian artery (LSA), while simultaneously excluding the diseased aortic segment by deploying additional stent grafts parallel to the main aortic component. These methods include chimney, snorkel and periscope techniques, which differ in the orientation of the branch grafts, depending on the target vessel and corresponding access.5 Before the procedure, careful planning is required using high-resolution computed tomography angiography to determine the most appropriate approach and the appropriate device sizes. Usually, a 20-30% oversizing is considered with respect to the main aortic component. Typically, the femoral artery is the chosen access for deployment of the primary thoracic stent graft, while carotid and/or axillary accesses are necessary for the deployment of parallel stents. During the procedure, the main aortic component is usually deployed first. At the same time, the parallel grafts are already correctly positioned through the sheaths (1 to 2 cm beyond the proximal margin of the primary graft). Grafts’ overlapping length should be at least 5 cm.6

One of the key advantages of PGTs is their immediate availability, which makes them particularly valuable in emergent situations, such as ruptured aneurysms or complicated dissections, where timely intervention can be lifesaving. Additionally, PGTs are less expensive and offer greater flexibility and versatility in addressing complex cases with challenging anatomy. A potential drawback associated with this technique is the increased risk of gutter-related endoleaks, which raises concern about long-term durability.5

Technical considerations of custom-made graft techniques

CMDs, specifically branched and fenestrated endografts, provide anatomically precise solutions for endovascular aortic arch repair. These devices are tailored using detailed preoperative imaging to match each patient's unique anatomy. On the one hand, branched endografts are deployed via femoral access for the main component, with additional carotid and/or axillary access to cannulate the arch branches. Subsequently, bridging stents are used to connect the supra-aortic trunks to the internal branches of the main graft. Regarding these components, a variety of stents (balloon- or self-expandable) have been used, depending on the indications for use (IFUs) and the individual surgeon’s preference. On the other hand, fenestrated endografts incorporate pre-designed fenestrations that should be aligned with the origins of the supra-aortic arteries, as planned in advance.

The manufacturing process for both branched and fenestrated devices is meticulous, and these devices are theoretically associated with better sealing and longer durability compared with off-the-shelf solutions, thereby minimising the risk of endoleaks and stent migration. However, these procedures are usually more complex, with steeper learning curves and are time-consuming, making them less suitable for emergent cases.7

Descriptive analysis according to the outcomes of interest

A total of 38 articles comprising 871 patients were included in this analysis (14 studies with 214 patients treated with PGTs and 24 studies with 657 patients treated with CMDs). The mean age was 72.1±10.4 years in the CMD group and 65.7±12.3 years in the PGT group. The majority of patients were male in both groups. The most frequently reported aetiologies were degenerative aneurysms, followed by aortic dissections and post-dissection aneurysms. Within the PGT group, the proportions of degenerative aneurysms and aortic dissections were relatively balanced, whereas CMDs were predominantly used for degenerative aneurysms. The remaining indications for treatment included pseudoaneurysms, penetrating aortic ulcers and type 1a endoleaks after previous TEVAR, Tables 1 and 2.8-45

In the clinical setting, the proportion of urgent cases within PGT series varied widely, ranging from exclusively urgent to predominantly elective. By contrast, CMD procedures were primarily elective because of manufacturing time constraints. However, some series describe the use of CMDs in urgent cases where stent grafts were either already available for the patient or sourced from another patient with similar anatomy.26 Tables 3-6 present the incidence of events for each study’s outcomes of interest.

Table 1 Details of studies reporting outcomes of parallel graft techniques for total endovascular aortic arch repair, included in this review 

Authors Year Study type N Age (years) ± SD Male n (%) Etiology n (%) LSA preservation n (%)
Voskresensky et al.8 2017 R 5 74.8 ± 10.4 4 (80) 3 (60) Degenerative Aneurysm 1(20) EL Ia after TEVAR 1 (20) PAU 4 (80)
Wang et al.9 2017 R 22 61.5 ± 7.7 20 (90.9) 19 (86.4) Aortic Dissection 2 (9.1) Degenerative Aneurysm 1 (4.5) EL Ia after TEVAR 1 (4.5)
Shahverdyan et al.10 2017 R 19 71.5 ± 8.5 17 (89.4) 6 (31.6) Degenerative Aneurysm 4 (21.1) Aortic Dissection 5 (26.3) EL Ia after TEVAR 3 (15.8) PAU 1 (5.3) Pseudoaneurysm NR
Wang et al. 11 2018 R 7 56.1 ± 10.8 5 (71.4) 4 (57.1) Degenerative Aneurysm 2 (28.6) Aortic Dissection 1 (14.3) EL Ia after TEVAR 7 (100)
Kanaoka et al. 12 2020 R 22 77.0 ± 10.0 15 (68.2) 22 (100) Degenerative Aneurysm 3 (13.6)
Teymouri et al.13 2023 R 5 63.8 ± 6.0 3 (60) 4 (80) Aneurysm + Dissection 1 (20) Degenerative Aneurysm 5 (100)
Gao et al.14 2017 R 10 55.6 ± 9.2 8 (80) 10 (100) Aortic Dissection 5 (50)
Olivas-Flores et al.15 2019 R 6 52.5 ± 12.4 5 (83.3) 6 (100) Aortic Dissection 0 (0)
Guo et al.16 2020 R 11 60.5 ± 11.9 9 (81.8) 4 (36.4) Pseudoaneurysm 3 (27.3) Aortic Dissection 3 (27.3) Degenerative Aneurysm 1 (9.1) EL Ia after TEVAR 11 (100)
Bosiers et al.17 2016 R 13 NR NR NR NR
Pecoraro et al.18 2017 R 20 NR NR NR NR
Bao et al.19 2021 R 6 NR NR NR 6 (100)
Zhao et al.20 2019 R 30 NR NR NR NR
Ahmad et al.21 2021 R 38 NR NR NR NR

SD: Standard deviation; LSA: Left subclavian artery; R: Retrospective; P: Prospective; TEVAR: Thoracic endovascular aortic repair; EL: Endoleak; PAU: Penetrating aortic ulcer; NR: Not reported

Table 2 Details of studies reporting outcomes of custom-made stent grafts for total endovascular aortic arch repair, included in this review 

Authors Year Study type N Age (years) ± SD Male n (%) Etiology n (%) LSA preservation n (%)
Kurimoto et al.22 2015 R 37 78.2 ± 6.6 29 (78.4) 31 (83.8) Degenerative Aneurysms 3 (8.1) Aortic dissections 2 (5.6) Aneurysm + Dissection 1 (2.7) Pseudoaneurysm 26 (70.3)
Spear et al.23 2015 R 27 74 (range 69-77) 22 (81.5) 19 (70.4) Degenerative Aneurysm 8 (29.6) Post-dissection Aneurysm 27(100)
Tazaki et al.24 2017 R 7 NR NR 7 (100) Degenerative Aneurysm 7 (100)
Spear et al.25 2017 R 18 64.0 (range 60-73) 15 (83.3) 18 (100) Post-dissection Aneurysm 11 (57.9)
Law et al.26 2019 R 11 67.0 ± 14.0 5 (45) 6 (54.5) Degenerative Aneurysm 3 (27.3) Aortic Dissection 1 (9.1) Pseudoaneurysm 11 (100)
Tsilimparis et al.27 2019 R 54 68.0 ± 10.0 38 (70.4) 26 (48.1) Post-dissection Aneurysm 24 (44.4) Degenerative Aneurysm 4 (7.4) PAU 54 (100)
Ferrer et al.28 2019 R 24 75.2 ± 6.5 24 (100) 13 (4.2) Degenerative Aneurysm 9 (37.5) PAU 2 (8.3) Aortic Dissection 23 (95.8)
Verscheure et al.29 2019 R 70 69.1 (range 62.4-74.1) 70 (100) Aortic Dissection 70 (100)
Weijde et al.30 2020 R 11 73.8 ± 4.8 9 (81.8) 9 (81.8) Degenerative Aneurysm 2 (18.2) Pseudoaneurysm 8 (72.7)
Tsilimparis et al.31 2019 R 12 NR NR 10 (83.3) Degenerative Aneurysm 2 (16.7) Post-dissection Aneurysm 10 (83.3)
Tenorio et al.32 2021 R 39 70.0 ± 7.0 31 (79.5) 25 (64.1) Post-dissection Aneurysm 14 (35.9) Degenerative Aneurysm 39 (100)
Czerny et al.33 2021 R 43 73.0 ± 9.0 33 (76.7) 26 (61) Degenerative Aneurysm 8 (19) PAU 7 (16) Post-dissection Aneurysm 2 (5) Unknown 34 (79.1)
Li et al.34 2021 R 16 54.5 ± 11.3 16 (100) 16 (100) Aortic Dissection 16 (100)
Hauck et al.35 2022 R 54 77.4 ± 8.1 44 (81.5) 40 (74.1) Degenerative Aneurysm 7 (13) Aortic Dissection 7 (13) PAU 33 (61.1)
Lee et al.36 2024 R 5 78.0 ± 13.0 4 (80) NR 5 (100)
Zhang et al.37 2023 R 15 68 (range 64-73) 14 (93.3) 10 (66.7) Degenerative Aneurysms 3 (13.3) Post-dissection Aneurysm 3 (13.3) PAU 1 (6.7) Pseudoaneurysm NR
Abisi et al.38 2023 P 18 67.8 ± 11.0 15 (83.3) 10 (55.6) Degenerative Aneurysm 6 (33.3) Post-dissection Aneurysms 2 (11.1) PAU
Nana et al.39 2023 R 8 72.3 ± 27.0 7 (87.5) 4 (50) Aortic Dissection 3 (37.5) Degenerative Aneurysm 1 (12.5) Pseudoaneurysm 8 (100)
Becker et al.40 2023 R 10 75.0 ± 11.7 6 (60) 10 (100) Pseudoaneurysm 8 (80)
Iglesias et al.41 2023 R 12 74.0 ± 7.0 12 (100) 9 (75) Degenerative Aneurysm 1 (8.3) Pseudoaneurysm 1 (8.3) PAU 1 (8.3) EL Ia after TEVAR 12 (100)
Knapsis et al.42 2024 R 6 74.2 ± 4.1 5 (83.3) 4 (66.7) Post-dissection Aneurysms 2 (33.3) PAU 6 (100)
Becker et al.43 2024 R 22 72 (range 63-79) 16 (72.7) 13 (59.1) Post-dissection Aneurysm 9 (40.9) Degenerative Aneurysm 22 (100)
Fukushima et al.44 2024 R 30 75.9 ± 6.3 NR 30 (100) Degenerative Aneurysm 6 (20)
Jubouri et al.45 2023 R 108 70.7 ± 9.9 72 (66.6) 81 (75) Degenerative Aneurysm 27 (25) Aortic Dissection NR

SD: Standard deviation; LSA: Left subclavian artery; R: Retrospective; P: Prospective; TEVAR: Thoracic endovascular aortic repair; EL: Endoleak; PAU: Penetrating aortic ulcer; NR: Not reported

Table 3 30-day mortality, stroke/TIA, reintervention, aortic-related mortality and overall mortality of studies reporting parallel graft techniques for total endovascular aortic arch repair, included in this review 

Authors 30 day mortality n (%) Stroke/TIA n (%) Reintervention n (%) Aorta-related deaths n (%) Overall mortality n (%)
Voskresensky et al.8 0 (0) 1 (20) 0 (0) 1 (11) 0 (0)
Wang et al.9 0 (0) 2 (9.1) 2 (9.1) 0 (0) 0 (0)
Shahverdyan et al.10 NR NR NR NR NR
Wang et al.11 0 (0) 0 (0) 0 (0) 0 (0) 1 (14.3)
Kanaoka et al.12 1 (4.5) 1 (4.5) NR 1 (4.5) 4 (18.2)
Teymouri et al.13 0 (0) 0 (0) 2 (40) 0 (0) 1 (20)
Gao et al.14 2 (20) 0 (0) 1 (10) 0 (0) 0 (0)
Olivas-Flores et al.15 0 (0) 0 (0) 2 (33.3) 0 1 (16.7)
Guo et al.16 0 (0) 5 (45.5) 2 (18.2) 1 (9.1) 3 (27.3)
Bosiers et al.17 1 (7.7) NR NR 0 (0) NR
Pecoraro et al.18 2 (10) NR NR NR NR
Bao et al.19 NR 0 (0) NR NR NR
Zhao et al.20 NR NR NR NR NR
Ahmad et al.21 NR 5 (13.2) NR NR NR

TIA: Transient ischemic attack; NR: Not reported

Table 4 30-day mortality, stroke/TIA, reintervention, aortic-related mortality and overall mortality of studies reporting custom-made stent grafts for total endovascular aortic arch repair, included in this review 

Authors 30 day mortality n (%) Stroke/TIA n (%) Reintervention n (%) Aorta-related deaths n (%) Overall mortality n (%)
Kurimoto et al.22 0 (0) 2 (5.4) 4 (10.8) 0 (0) 6 (16.2)
Spear et al.23 0 (0) 3 (11.1) 6 (22.2) 1 (3.7) 1 (3.7)
Tazaki et al.24 2 (28.6) 3 (42.9) NR NR NR
Spear et al.25 1 (5.3) 2 (10.5) 6 (33.3) 1 (5.3) 4 (21.1)
Law et al.26 1 (9) 1 (9) 5 (45.5) 0 (0) 3 (27.3)
Tsilimparis et al.27 3 (5.6) 6 (11.1) 12 (22.2) 1 (1.9) 7 (13.0)
Ferrer et al.28 4 (16.7) 7 (29.2) 4 (16.7) 2 (8.3) 5 (20.8)
Verscheure et al.29 2 (2.9) 7 (10) 32 (45.7) 1 (1.4) 10 (14.3)
Weijde et al.30 2 (18.2) 4 (36.4) 2 (18.2) 1 (9.1) 2 (18.2)
Tsilimparis et al.31 1 (8.3) 1 (8.3) NR NR NR
Tenorio et al.32 2 (5.1) 2 (5.1) 14 (35.9) 0 (0) 4 (10.3)
Czerny et al.33 4 (9.3) 11 (25.6) 4 (9.3) 1 (2.3) 15 (34.9)
Li et al.34 1 (6.3) 1 (6.3) 2 (12.5) 1 (6.3) 5 (31.3)
Hauck et al.35 2 (3.7) 5 (9.3) 8 (14.8) 2 (3.7) 21 (38.9)
Lee et al.36 0 (0) 0 (0) NR 0 (0) 0 (0)
Zhang et al.37 1 (6.7) 2 (13.3) NR NR NR
Abisi et al.38 0 (0) 0 (0) 3 (16.7) 0 (0) 0 (0)
Nana et al.39 0 (0) 0 (0) 3 (37.5) 0 (0) 0 (0)
Becker et al.40 2 (20) 2 (20) 3 (30) 0 (0) 3 (30)
Iglesias et al.41 1 (8.3) 5 (41.7) 1 (8.3) 0 (0) 2 (16.7)
Knapsis et al.42 0 (0) 0 (0) 0 (0) NR NR
Becker et al.43 1 (4.5) 3 (13.6) 1 (4.5) 0 (0) 1 (4.5)
Fukushima et al.44 0 (0) 1 (3.3) 1 (3.3) 0 (0) 0 (0)
Jubouri et al.45 4 (3.7) 32 (29.6) 50 (46.3) 0 (0) 4 (3.7)

TIA: Transient ischemic attack; NR: Not reported

Table 5 Technical success, type I/III endoleaks, spinal chord ischaemia, branch patency, and duration of follow-up of studies reporting parallel graft techniques for total endovascular aortic arch repair, included in this review 

Authors Technical success n (%) Type I/III Endoleak n (%) Spinal cord ischemia n (%) Branch patency (%) Follow-up (months)
Voskresensky et al.8 4 (80) 1 (20) 0 (0) 100% NR
Wang et al.9 19 (86.3) 3 (13.6) NR 100% 28.0 ± 19.8
Shahverdyan et al.10 NR NR NR 100% NR
Wang et al.11 NR 3 (42.9) 0 (0) 100% 15.7 (range 9-20)
Kanaoka et al.12 18 (81.8) 6 (27.3) NR 95.5% NR
Teymouri et al.13 5 (100) 2 (40) NR 100% 54.0 ± 37.2
Gao et al.14 10 (100) 0 (0) NR 100% 17.0 ± 14.5
Olivas-Flores et al.15 6 (100) 1 (16.7) 0 (0) 100% 12.0
Guo et al.16 5 (45.5) 5 (45.5) 0 (0) 90.9% 27.4 ± 16.5
Bosiers et al.17 NR NR NR NR NR
Pecoraro et al.18 NR NR NR 100% NR
Bao et al.19 6 (100) NR NR 100% NR
Zhao et al.20 NR 3 (10) NR 100% NR
Ahmad et al.21 NR 8 (21.1) NR NR NR

ICU: Intensive care unit; IQR: Interquartile range; NR: Not reported

Table 6 Technical success, type I/III endoleaks, spinal chord ischaemia, branch patency, and duration of follow-up of studies reporting custom-made stent grafts for total endovascular aortic arch repair, included in this review 

Authors Technical success n (%) Type I/III Endoleak n (%) Spinal cord ischemia n (%) Patency (%) Follow-up (months)
Kurimoto et al.22 37 (100) 12 (32.4) 3 (8.1) 100% 16.9 ± 12.8
Spear et al.23 27 (100) 0 (0) 2 (7.4) 100% 12 (IQR 6-12)
Tazaki et al.24 7 (100) NR 0 (0) 100% NR
Spear et al.25 16 (84.2) 1 (5.3) 0 (0) 100% NR
Law et al.26 11 (100) 0 (0) 0 (0) 100% 6 (range 1-28)
Tsilimparis et al.27 53 (98.1) 2 (3.7) 3 (5.6) 98.1% 12.0 ± 9.0
Ferrer et al.28 23 (95.8) 0 (0) 0 (0) 100% 18 (range 1-60)
Verscheure et al.29 66 (94.3) 4 (5.7) 0 (0) 100% 10 (IQR 4.6-21.4)
Weijde et al.30 11 (100) 1 (9.1) 1 (9.1) 90.9% 17 (range 3-42)
Tsilimparis et al.31 NR NR 1 (8.3) NR NR
Tenorio et al.32 39 (100) 5 (12.8) 0 (0) 100% 3.2 (IQR 1-14)
Czerny et al.33 NR 4 (9.3) 1 (2.3) NR 16.0 ± 18.0
Li et al.34 16 (100) 1 (6.3) NR 87.5% 98 (range 0-119)
Hauck et al.35 53 (98.1) 2 (3.7) 1 (1.9) 98.1% 35.8 ± 38.2
Lee et al.36 5 (100) NR 0 (0) 100% 13.1
Zhang et al.37 15 (100) NR 1 (6.7) NR 12.3 (IQR 6.3-30.6)
Abisi et al.38 18 (100) 4 (22.2) 0 (0) 100% 9 (range 4-18)
Nana et al.39 8 (100) 1 (12.5) 0 (0) 100% 2 (range 1-4)
Becker et al.40 9 (90) 3 (30) 0 (0) 100% 20 (IQR 55)
Iglesias et al.41 12 (100) 1 (8.3) 0 (0) 100% 15.5 (range 0-44)
Knapsis et al.42 6 (100) 0 (0) 0 (0) 100% NR
Becker et al.43 12 (100) 4 (18.2) 0 (0) 100% NR
Fukushima et al.44 30 (100) 0 (0) 1 (3.3) 100% 34.9 (range 1-73)
Jubouri et al.45 NR NR NR 74% NR

ICU: Intensive care unit; IQR: Interquartile range; NR: Not reported

Technical success

Technical success rates were high in both groups, mostly ranging from 80% to 100%. There was only one outlier study reporting a lower technical success rate of 45.5% in PGT group.16 CMDs exhibited the most consistent success rates, typically exceeding 95%, owing to their patient-specific design and precise deployment.

Thirty-day, aorta-related and overall mortality

No clear superiority was observed between the two approaches, though the overall mortality rate appeared slightly higher in the CMD group, ranging from 0 to 38.9%22-45 (versus 0-27.3% in the PGT group).8-21 The thirty-day mortality rate ranged from 0 to 20% in PGTs8-21 and from 0 to 28.6% in CMDs.22-45 Aorta-related mortality rates were also low, ranging from 0 to 11% in PGTs8-21 and from 0 to 9.1% in CMDs.22-45

Stroke rate and neurological complications

Stroke rate appeared higher in the CMD group, ranging from 0 to 42.9%.22-45 By contrast, only one study on PGTs reported a stroke rate of 45.5%,16 whereas the remaining PGT studies reported stroke rates <20%. Moreover, although LSA preservation was substantially more common with CMDs, no spinal cord ischaemia (SCI) cases were reported in the PGT group, whereas CMD had an SCI rate ranging from 0 to 9.1%.22-45

Occurrence of endoleaks and reintervention rates

As expected, the occurrence of type I and III endoleaks was higher in the PGT group, ranging from 0 to 45.5%8-21 (versus 0-32.4% in the CMD group).22-45 Nevertheless, this discrepancy did not translate into a clear difference in reintervention rates. In fact, the overall reintervention rate appeared higher in CMD patients, whereas aorta-related reinterventions remained similar between the two groups. Non-aorta-related reinterventions in the CMD group were mostly access complications.

Target vessel patency

The mean follow-up was 28.3±22.1 months for PGTs and 20.7±25.2 months for CMDs. During this period, high target vessel patency rates were observed in both groups, typically exceeding 95%.8-45

Discussion

This comprehensive review compared the outcomes of PGTs and CMDs in total endovascular aortic arch repair, focusing on technical aspects, clinical outcomes, and strengths and limitations. Our findings suggest that although both strategies achieved acceptable mid-term results, they have distinct advantages and drawbacks that must be carefully weighed in clinical decision-making. Both techniques demonstrated comparable 30-day mortality and reasonably low aortic mortality, reinforcing the safety of these endovascular strategies for treating complex aortic arch pathologies. Long-term durability was further supported by high target vessel patency, generally exceeding 95% in both groups.

Conversely, the overall mortality and stroke rates seemed to be slightly higher in CMDs, possibly due to the older age of this group or the aortic arch manipulation during the deployment of branched/fenestrated endografts in zone 0 TEVAR. As expected, PGTs were associated with lower technical success rates and a higher incidence of type I and III endoleaks, probably through the gutters of the stent grafts. However, aorta-related reintervention rates appeared comparable between the two groups, which may reflect different failure patterns - PGTs may be associated with more endoleaks that remain asymptomatic and are treated conservatively, whereas CMDs may require secondary interventions for device-related issues or disease progression. Although LSA preservation was substantially more common with CMDs, SCI was reported in nine CMD studies but in none of the PGT studies, which is probably associated with differences in the extent of aortic coverage.

Study limitations

The absence of direct comparative data between PGTs and CMDs, together with the retrospective design of the included studies, is a major limitation of this review. Additionally, the lack of outcome stratification by the proximal landing zone, combined with the use of single-chimney techniques with extra-anatomical bypasses (rather than endovascular revascularisation of both BCT and LCCA), led to the exclusion of several potentially relevant studies from our analysis. Furthermore, time-stratified data were not available in the included studies, preventing assessment of learning-curve effects.

Conclusion

PGTs appear to have non-inferior outcomes compared with CMDs in terms of stroke rate and in-hospital and overall mortality. Although the occurrence of type I and III endoleaks remains higher in PGTs, this does not seem to affect aorta-related reintervention rates. Thus, PGTs should be considered a viable option for treating complex aortic arch pathology, particularly in fragile patients, those with adverse anatomy, or in emergent situations. The lack of high-quality data underscores the need for prospective, comparative studies to better define the optimal approach for total aortic arch reconstruction.

Acknowledgements

None

Conflicts of interest

None

Funding

None

Data availability

By request to authors

Ethics Approval

Not applicable

Informed Consent

No written informed consent was required due to the study design

Declaration of Generative AI and AI-Assisted Technologies in the Writing Process

AI-assisted translation to British English language was utilised. No AI generative technology was applied in the writing or editing of scientific content.

References

1. Czerny M, Schmidli J, Adler S, van den Berg JC, Bertoglio L, Carrel T, et al. Current options and recommendations for the treatment of thoracic aortic pathologies involving the aortic arch: an expert consensus document of the European Association for Cardio-Thoracic surgery (EACTS) and the European Society for Vascular Surgery (ESVS). Eur J Cardiothorac Surg. 2019;55:133-62. [ Links ]

2. Brown JA, Arnaoutakis GJ, Szeto WY, Serna-Gallegos D, Sultan I. Endovascular repair of the aortic arch: State of the art. J Card Surg. 2021;36:4292-300. [ Links ]

3. Ahmad W, Wegner M, Dorweiler B. Meta-analysis and meta-regression of the total endovascular aortic repair in aortic arch. Vasa. 2023;52:175-85. [ Links ]

4. Zhu L, Li X, Lu Q. A systematic review and meta-analysis of thoracic endovascular aortic repair with the proximal landing zone 0. Front Cardiovasc Med. 2023;10:1034354. [ Links ]

5. Li Y, Hu Z, Wang J, Zhang Y, Chen Z, Zhang H. Endovascular Chimney Technique for Aortic Arch Pathologies Treatment: A Systematic Review and Meta-Analysis. Ann Vasc Surg. 2018;47:305-15. [ Links ]

6. Kwiecinski J, Cheng CP, Uberoi R, Hadi M, Hempel P, Degel C, You Z. Thoracic aortic parallel stent-graft behaviour when subjected to radial loading. J Mech Behav Biomed Mater. 2021;118:104407. [ Links ]

7. Spath P, Campana F, Tsilimparis N, Gallitto E, Pini R, Faggioli G, et al. Outcomes of Fenestrated and Branched Endografts for Partial and Total Endovascular Repair of the Aortic Arch - A Systematic Review and Meta-Analysis. Eur J Vasc Endovasc Surg. 2024;67:106-16. [ Links ]

8. Voskresensky I, Scali ST, Feezor RJ, Fatima J, Giles KA, Tricarico R, et al. Outcomes of thoracic endovascular aortic repair using aortic arch chimney stents in high-risk patients. J Vasc Surg. 2017;66:9-20. [ Links ]

9. Wang T, Shu C, Li QM, Li M, Li X, He H, et al. First experience with the double chimney technique in the treatment of aortic arch disease. J Vasc Surg. 2017;66:1018-27. [ Links ]

10. Shahverdyan R, Mylonas S, Gawenda M, Brunkwall J. Single-center mid-term experience with chimney-graft technique for the preservation of flow to the supra-aortic branches. Vascular. 2018;26:175-82. [ Links ]

11. Wang L, Huang Y, Guo D, Xu X, Chen B, Jiang J, et al. Application of triple-chimney technique using C-TAG and Viabahn or Excluder iliac extension in TEVAR treatment of aortic arch dilation diseases. J Thorac Dis. 2018;10:3783-90. [ Links ]

12. Kanaoka Y, Ohki T, Maeda K, Shukuzawa K, Baba T, Tezuka M, et al. Outcomes of Chimney Thoracic Endovascular Aortic Repair for an Aortic Arch Aneurysm. Ann Vasc Surg. 2020;66:212-9. [ Links ]

13. Teymouri M, Mehta M, Paty P, Jayakumar L, Kostun ZW. Complete zone 0-10 aortic endovascular reconstruction. J Vasc Surg Cases Innov Tech. 2023;9:101111. [ Links ]

14. Gao F, Zeng Q, Lin F, Ge X. Sandwich Technique for Endovascular Repair of Acute Type A Aortic Dissection. J Endovasc Ther. 2017;24:647-53. [ Links ]

15. Olivas-Flores R, Llamas-Macias FJ, Nuño-Escobar C, Calderón-Llamas MA, Landín-Alcaraz A, Cota-Gallegos I, et al. Reparación endovascular de aorta torácica con técnica de doble chimenea: una alternativa a la cirugía de arco aórtico posterior a la realización de Bentall-Bono en la disección aórtica tipo A. Rev Mex Angiol. 2019;47:36-42. [ Links ]

16. Guo B, Guo D, Chen B, Shi Z, Dong Z, Yan C, et al. Endovascular Outcomes in Aortic Arch Repair with Double and Triple Parallel Stent Grafts. J Vasc Interv Radiol. 2020;31:1984-92. [ Links ]

17. Bosiers MJ, Donas KP, Mangialardi N, Torsello G, Riambau V, Criado FJ, et al. European Multicenter Registry for the Performance of the Chimney/Snorkel Technique in the Treatment of Aortic Arch Pathologic Conditions. Ann Thorac Surg. 2016,101:2224-30. [ Links ]

18. Pecoraro F, Lachat M, Cayne NS, Pakeliani D, Rancic Z, Puippe G, et al. Mid-term Results of Chimney and Periscope Grafts in Supra-aortic Branches in High Risk Patients. Eur J Vasc Endovasc Surg. 2017;54:295-302. [ Links ]

19. Bao X, Zhao Y, Li T, Wu M, Zeng Z, Gao M, et al. Chimney Graft Technique Combined With Embolization for Treating Ruptured Aortic Arch Lesions. Front Cardiovasc Med. 2021;8:711283. [ Links ]

20. Zhao Y, Feng J, Yan X, Zhu G, Zhou J, Li Z, et al. Outcomes of the Chimney Technique for Endovascular Repair of Aortic Dissection Involving the Arch Branches. Ann Vasc Surg. 2019;58:238-47. [ Links ]

21. Ahmad W, Liakopoulos OJ, Mylonas S, Wegner M, Brunkwall J, Dorweiler B. Long-Term Outcomes after Thoracic Endovascular Aortic Repair Using Chimney Grafts for Aortic Arch Pathologies: 10 Years of Single-Center Experience. Ann Vasc Surg. 2021;72:400-8. [ Links ]

22. Kurimoto Y, Maruyama R, Ujihira K, Nishioka N, Hasegawa K, Iba Y, et al. Thoracic Endovascular Aortic Repair for Challenging Aortic Arch Diseases Using Fenestrated Stent Grafts From Zone 0. Ann Thorac Surg. 2015;100:24-32. [ Links ]

23. Spear R, Haulon S, Ohki T, Tsilimparis N, Kanaoka Y, Milne CP, et al. Editor&apos;s Choice - Subsequent Results for Arch Aneurysm Repair with Inner Branched Endografts. Eur J Vasc Endovasc Surg. 2016;51:380-5. [ Links ]

24. Tazaki J, Inoue K, Higami H, Higashitani N, Toma M, Saito N, et al. Thoracic endovascular aortic repair with branched Inoue Stent Graft for arch aortic aneurysms. J Vasc Surg. 2017;66:1340-8. [ Links ]

25. Spear R, Hertault A, Van Calster K, Settembre N, Delloye M, Azzaoui R, et al. Complex endovascular repair of postdissection arch and thoracoabdominal aneurysms. J Vasc Surg. 2018;67:685-93. [ Links ]

26. Law Y, Kölbel T, Detter C, Rohlffs F, von Kodolitsch Y, Makaloski V, et al. Emergency Use of Branched Thoracic Endovascular Repair in the Treatment of Aortic Arch Pathologies. Ann Thorac Surg. 2019;107:1799-806. [ Links ]

27. Tsilimparis N, Detter C, Law Y, Rohlffs F, Heidemann F, Brickwedel J, et al. Single-center experience with an inner branched arch endograft. J Vasc Surg. 2019;69:977-85. [ Links ]

28. Ferrer C, Cao P, Coscarella C, Ferri M, Lovato L, Camparini S, et al. iTalian RegIstry of doUble inner branch stent graft for arch PatHology (the TRIUmPH Registry). J Vasc Surg. 2019;70:672-82. [ Links ]

29. Verscheure D, Haulon S, Tsilimparis N, Resch T, Wanhainen A, Mani K, et al. Endovascular Treatment of Post Type A Chronic Aortic Arch Dissection With a Branched Endograft: Early Results From a Retrospective International Multicenter Study. Ann Surg. 2021;273:997-1003. [ Links ]

30. Weijde E, Heijmen RH, van Schaik PM, Hazenberg CEVB, van Herwaarden JA. Total Endovascular Repair of the Aortic Arch: Initial Experience in the Netherlands. Ann Thorac Surg. 2020;109:1858-63. [ Links ]

31. Tsilimparis N, Law Y, Rohlffs F, Spanos K, Debus ES, Kölbel T. Fenestrated endovascular repair for diseases involving the aortic arch. J Vasc Surg. 2020;71:1464-71. [ Links ]

32. Tenorio ER, Oderich GS, Kölbel T, Dias NV, Sonesson B, Karelis A, et al. Multicenter global early feasibility study to evaluate total endovascular arch repair using three-vessel inner branch stent-grafts for aneurysms and dissections. J Vasc Surg. 2021;74:1055-65 [ Links ]

33. Czerny M, Berger T, Kondov S, Siepe M, Saint Lebes B, Mokrane F, et al. Results of endovascular aortic arch repair using the Relay Branch system. Eur J Cardiothorac Surg. 2021;60:662-8. [ Links ]

34. Li X, Zhang L, Song C, Zhang H, Xia S, Li H, et al. Long-Term Outcomes of Thoracic Endovascular Repair for Aortic Arch Dissection Using Customized Single-Branched Fenestrated Stent-Graft. Vasc Endovascular Surg. 2021;55:577-85. [ Links ]

35. Hauck SR, Kupferthaler A, Kern M, Rousseau H, Ferrer C, Iwakoshi S, et al. Branched versus fenestrated thoracic endovascular aortic repair in the aortic arch: A multicenter comparison. J Thorac Cardiovasc Surg. 2022;164:1379-89. [ Links ]

35. Lee KB, Porras-Colon J, Scott CK, Chamseddin K, Baig MS, Timaran CH. Early Results and Feasibility of Total Endovascular Aortic Arch Repair Using 3-Vessel Company-Manufactured and Physician-Modified Stent-Grafts. J Endovasc Ther. 2024;31:1197-207. [ Links ]

37. Zhang H, Rong D, Liu F, Ge Y, Wei R, Zhu Y, et al. Case series of total endovascular repair of the aortic arch with a modular inner-branched stent-graft system: first-in-man experience. Br J Surg. 2023;110:1084-6. [ Links ]

38. Abisi S, Elnemr M, Clough R, Alotaibi M, Gkoutzios P, Modarai B, et al. The Development of Totally Percutaneous Aortic Arch Repair With Inner-Branch Endografts: Experience From 2 Centers. J Endovasc Ther. 2025;32:730-8. [ Links ]

39. Nana P, Houérou TL, Guihaire J, Gaudin A, Fabre D, Haulon S. Early Outcomes on Triple-Branch Arch Device With Retrograde Left Common Carotid Branch: A Case Series. J Endovasc Ther. 2025;32:802-12. [ Links ]

40. Becker D, Stana J, Prendes CF, Konstantinou N, Öz T, Pichlmaier M, et al. Endovascular arch repair of anastomotic aneurysm and pseudoaneurysm in patients after open repair of the ascending aorta and aortic arch: a case series. Eur J Cardiothorac Surg. 2023;64:345. [ Links ]

41. Iglesias Iglesias C, González Canga C, Abril Ramiro F, Alonso Pastor A, Álvarez Marcos F, Alonso Pérez M. An early single-center experience with the Relay double inner-branch arch endograft. J Thorac Dis. 2023;15:6721-9. [ Links ]

42. Knapsis A, Seker MM, Schelzig H, Wagenhäuser MU. Comparative Evaluation of the Short-Term Outcome of Different Endovascular Aortic Arch Procedures. J Clin Med. 2024;13:4594. [ Links ]

43. Becker D, Stana J, Prendes C, Ali A, Pichlmaier M, Peterss S, et al. The Use of Short Dilator Tip in Endovascular Branched Arch Repair: A Case Series. J Endovasc Ther. 2024;18:15266028241283713. [ Links ]

44. Fukushima S, Ohki T, Tachihara H, Shukuzawa K, Ohmori M, Ozawa H, et al. Clinical impact of proximal fixation augmentation using the Najuta thoracic fenestrated stent graft during endovascular treatment for distal aortic arch aneurysm. J Vasc Surg. 2024;80:949-56. [ Links ]

45. Jubouri M, Surkhi AO, Al-Tawil M, Piffaretti G. Driving Technology for Thoracic Endovascular Aortic Repair: An International Analysis of Single-versus Double-Branch RELAY Outcomes. Ann Vasc Surg. 2023 Aug;94:80-91. [ Links ]

Received: April 23, 2025; Accepted: November 11, 2025

Corresponding Author: Patrícia Carvalho | patricia.sofia.2407@gmail.com Department of Angiology and Vascular Surgery of the ULS Gaia/Espinho Rua Conceição Fernandes - Vilar de Andorinho, 4434-502, Vila Nova de Gaia

Creative Commons License This is an open-access article distributed under the terms of the Creative Commons Attribution License